Epsilon Chemical Test Procedure
Purpose
The purpose of this test is to perform a final examination of the epsilon before going to the customer, and to provide the customer with typical output and data. From these outputs, the customer can verify that the instrument is working properly on arrival and can gain some experience in its operation.
Instrument Installation
The PC should be set up according to the instructions included with the PC. The epsilon should be plugged into the mains supply (N.B. first check that the fuse panel has been set to the correct voltage). The COMPUTER
port of the epsilon is connected to a COM
port of the PC. If available, a BASi cell stand should be connected to the CELL STAND
port of the epsilon
Test Solution
2 mM potassium ferricyanide with 1 M potassium nitrate in water.
Preparation of the Test Solution
- Weigh 16.5 mg potassium ferricyanide and place in a 25 mL volumetric flask.
- Weigh 2.53 g potassium nitrate and add to the same volumetric flask.
- Add about 20 mL deionized water to dissolve the potassium ferricyanide and potassium nitrate.
- Dilute to 25 mL with deionized water.
Cell (C3 Cell Stand)
Platinum (PTE) Working Electrode (Black lead)
Platinum Wire Auxiliary Electrode (Red lead)
Silver/Silver Chloride RE-5 Reference Electrode (White lead)
Add 10-15 mL of the ferricyanide solution to the cell vial and place in the cell holder (see Cell Stand instructions). Polish the PTE with 0.05 mm alumina following the polishing instructions provided in the polishing kit.
If the RE-5 electrode is new, carefully remove the yellow plastic sheath before use. In addition, there may be air bubbles inside the electrode next to the CoralPor™ frit; these must be displaced (by flicking the electrode). RE-5 electrodes must be stored in 3M sodium chloride when not in use.
Procedure
- Turn the Power switch on the epsilon to on.
- Open the epsilon software by clicking the epsilon icon. The CS Dialog
box will appear, and will show the progress of establishment of the communication between the epsilon and the PC. If the link is established, the CS Dialog
box will automatically disappear. If the link is not established, click the Retry
button. If the link is lost after being established, use Reconnect Epsilon in the Experiment
menu to reestablish the link.
- Click New
in the File
menu (or use the F2
key) to set up a new experiment. The list of available techniques is displayed (Fig 1) (this list can also be accessed using Select New Experiment
in the Experiment
menu). It should be noted that there are some techniques that are labeled as DEMO. This label indicates that this technique is NOT active on this particular epsilon. However, it is possible to load a data file for that technique to examine the parameters and the typical output. If the RUN
button is clicked when a DEMO data file is displayed in the active window, an error message will be shown. The technique list shown in Figure 1
is the list for the basic epsilon (i.e., pulse techniques, stripping techniques, and multi-channel amperometry are NOT available).
Figure 1. Technique list.
- Select Cyclic Voltammetry. The Change Parameters
dialog box will now be displayed. Enter the values shown in Figure 2. Note that Switching Potential 2
is not required since there are only 2 segments (Initial Potential
to Switching Potential 1
to Final Potential). Once these changes have been entered, an experiment using these parameters can be run by clicking the RUN
button in this dialog box. If an experiment is not to be run immediately, but these parameters are to be saved, then the Apply
button must be clicked before exiting the dialog box. If Exit
is clicked before Apply, any changes in the parameters will be lost. After exiting the dialog box, an experiment can be run using either Run
in the Experiment
menu, in the pop-up menu obtained by clicking the RIGHT mouse button (Figure 3), the RUN
icon on the Tool Bar, or the F5
key. This icon will change to STOP
during the experiment, and can be used to abort the experiment.
Figure 2. Change Parameters
dialog box for Cyclic Voltammetry.
Figure 3. Right mouse button pop-up menu.
- After the experiment has been run, the voltammogram will be displayed (Figure 4). Note the information about the experiment and the peak parameters on the right side of the graph (this can be removed by clicking Text Info (Right Column)
in the Graph-Display
menu or the pop-up menu).
Figure 4. Cyclic Voltammetry
graph.
- A specific area of the graphic can be enlarged by either using the mouse cursor (and the left mouse button) to define the area (Figure 5) or entering the x and y values in the Manual Zoom
dialog box (Figure 6). The original graph can be restored using Auto Zoom
in the Graph-Display
menu.
Figure 5.
Figure 6. Manual Zoom
dialog box.
- Use Save
in the File
menu (or the F4
key) to save the data in the active experiment window (it should be noted that the each technique has its own extension; for example, .cv0 for cyclic voltammetry). The data is saved in a binary format, but, once saved, it can be converted to a number of different text formats using Convert to Text File
in the File
menu (Figure 7). Select the file(s) to be converted, the format, and the delimiter, and then click Convert
to start the conversion.
Figure 7. Convert to Text
dialog box.
- Various experimental data can be entered into the Notes
in the Experiment
menu (Figure 8). These notes will also be saved when the experimental data is saved.
Figure 8. Notes
dialog box.
- Open the Change Parameters
dialog box (using Change Parameters
in either the Experiment
menu, the pop-up menu, or the F6
key), and change the Scan Rate
to 200 mV/s. Run the experiment again. Note that the new data is displayed in the experiment window; that is, the new experimental data has replaced the old experimental data. Save this data. Change the scan rate to 500 mV/s, run the experiment, and save the data.
- The three data sets run at different scan rates can be displayed on the same sets of axes using the File Overlay
function in the Graph-Display
menu. The first time this function is selected after the software is started, an empty set of axes is displayed. Subsequent activations will display the previously selected data sets. Click the right mouse, and select Setup
from the pop-up menu. The File Overlay Setup
dialog box will be displayed (Figure 9).
Figure 9. File Overlay Setup
dialog box.
- The scaling of the axes is determined by the Base File. Therefore, when comparing voltammograms with the same potential range, the voltammogram with the highest current should be selected as the Base File. In this example, the cyclic voltammogram run at a scan rate of 500 mV/s should be the Base File. Select the name of this data file, and click New. This data file is now displayed on the axis set (which remains in the background). The other two data files (cyclic voltammograms run at 100 and 200 mV/s) should now be selected (using standard Windows functions). Once selected, click Add
to display these data sets on the axis set. Click Exit
to remove the Setup
dialog box. The overlayed data sets can now be viewed (Figure 10).
Figure 10. Overlayed cyclic voltammograms.
- Click New
in the File
menu, and select Cyclic Voltammetry
again. A new experiment window will appear, together with an associated Change Parameters
dialog box. Enter the parameters shown in Figure 11, and run this experiment. Since there are 3 segments in this experiment, all 4 potential parameters must be defined (Initial Potential
to Switching Potential 1
to Switching Potential 2
to Final Potential).
Figure 11.
- When this experiment has finished, both
experiment windows are displayed in the main epsilon window (Figure 12). Up to 6 experiment windows can be displayed. However, only one experiment window can be active at any one time, and any epsilon functions and operations are applied only to the data in the active experiment window.
Figure 12.
- Click New, and select Chronoamperometry/Chronocoulometry
from the list of techniques. A third window will appear, together with an associated Change Parameters
dialog box. Enter the parameters shown in Figure 13. Running the experiment generates the plot shown in Figure 14.
Figure 13. Change Parameters dialog box for Chronoamperometry/Chronocoulometry.
Figure 14. Chronoamperometry/Chronocoulometry
graph.
- The data from a Chronoamperometry/Chronocoulometry
experiment can be plotted in a number of different formats, which can be selected using Select Graph
in the pop-up menu in the Chronoamperometry/Chronocoulometry
experiment window. The Q vs. sqrt(T)
plot is shown in Figure 15.
Figure 15. Q vs. sqrt(T)
plot for Chronoamperometry/Chronocoulometry.
- The epsilon software can calculate the slope and intercept of the linear Q vs sqrt(T) plot when it is displayed. Selecting Data - Calculate CA-SIR
from the pop-up menu generates the information box shown in Figure 16. The lines used for the linear fitting are the dashed blue lines in Figure 15
(note that the first 20% of the data points are not used in the calculation, due to interference from the charging current and other experimental artifacts).
Figure 16. Calculation of slope and intercept of linear plot.
- Click New, and select Chronopotentiometry. A fourth experiment window will be opened, together with the associated Change Parameters
dialog box. Enter the parameters shown in Figure 17, then run the experiment (note the sign convention for the current - cathodic (reduction) currents are positive for polarographic). Typical data for a Chronopotentiometry
experiment is shown in Figure 18.
Figure 17. Change Parameters
dialog box for Chronopotentiometry.
Figure 18. Chronopotentiometry
graph.
- Click New, and select Square Wave Voltammetry
(note that this requires the optional Basic Plus software). A fifth experiment window will be opened, together with the associated Change Parameters
dialog box. Enter the parameters shown in Figure 19, then run the experiment. Typical data for a Square Wave Voltammetry
experiment is shown in Figure 20.
Figure 19. Change Parameters
dialog box for Square Wave Voltammetry.
Figure 20. Default plot for Square Wave Voltammetry.
- The default plot for a Square Wave Voltammetry
experiment is the difference current; that is, the current on the forward cycle less the current on the reverse cycle. The forward and reverse currents are also available under Select Graph
in the pop-up menu in the Square Wave Voltammetry
experiment window. The forward current data set is shown in Figure 21.
Figure 21. Forward current plot for Square Wave Voltammetry.
This completes the chemical test. All printed output should be shipped with the instrument. The power cord and cell lead (with alligator clips on the cell end) are shipped with the epsilon.
Copyright © 2000 BASi E'CHEM